Biocompatibility and Toxicology Interview Questions for Medical Devices

Eight biocompatibility interview questions with model answers and three case studies on a supplier additive change, a cytotoxicity result and a predicate material claim.

Updated · AptSkill MedTech

Biocompatibility and Toxicology Interview Questions for Medical Devices: Regulatory Affairs Careers (AptSkill MedTech)

Biocompatibility work answers one question: will the materials in this device, in the form the patient meets, cause harm? Interviewers for biocompatibility, toxicology and biological safety roles want to see that you can run a risk-based biological evaluation, not just order a test battery, and that you can explain results to engineers, suppliers and regulators. This guide gives you 8 model questions with answers and three case studies with ideal answers.

What to prepare

Bring one device you evaluated or supported: its materials, the contact category and duration, what existing data you used, what you tested and why. Refresh the ISO 10993 biological evaluation guide, risk management and supplier controls. Check the current edition of ISO 10993-1 and FDA’s guidance on its use before the interview, because both are revised.

Eight common questions with model answers

1. Walk me through a biological evaluation for a new device.

Model answer: “I treat it as part of risk management under ISO 10993-1. I describe the device, its materials, processing aids and residues, then classify the body contact by nature and duration. I review existing data and chemical characterisation first, identify the endpoints that still need evidence, test those on the finished sterilised device, and record the rationale in a biological evaluation plan and report. A change to a material or process triggers a reassessment.”

2. How do contact category and duration drive the endpoints?

Model answer: “The category is surface, external communicating or implant, and the duration is limited (up to 24 hours), prolonged (over 24 hours up to 30 days) or long-term (over 30 days). Together they determine which endpoints to consider, for example cytotoxicity, sensitisation and irritation for most contact, and more systemic and implantation endpoints as contact becomes deeper and longer. The table is a starting point, and I justify any endpoint I leave out.”

3. What is chemical characterisation and why has it become central?

Model answer: “It identifies and quantifies what can be released from the device, using extraction conditions that match or exaggerate clinical use, under ISO 10993-18. A toxicologist then compares the exposure with tolerable limits under ISO 10993-17. It is central because good chemical data can show an acceptable margin of safety and reduce or replace some animal testing, and it makes material changes easier to assess.”

4. What does a toxicological risk assessment add?

Model answer: “It turns chemical data into a safety judgement. For each substance found, the toxicologist estimates patient exposure, compares it with a tolerable intake and states the margin. It handles substances with little test history and gives the evaluation report a defensible, quantitative basis.”

5. How do sterilisation and ageing affect your testing?

Model answer: “The patient meets the finished, sterilised, aged device, so that is what I test or extract, using the worst-case sterilisation exposure. Sterilisation can leave residues, such as ethylene oxide residuals controlled under ISO 10993-7, or change the material. Ageing and shelf life can change what leaches, so I confirm the sample condition matches the claim.”

6. When can you justify not testing an endpoint?

Model answer: “When existing evidence already covers it: the same material, processing and contact has a documented history of safe use, there is sufficient chemical and toxicological data, or the endpoint is not relevant for that contact. I write the justification in the plan, with references, because reviewers ask for the reasoning, not only the result.”

7. What changes trigger a biocompatibility reassessment?

Model answer: “A new or changed material, supplier, additive, colourant, adhesive, coating, cleaning agent, process parameter or sterilisation method, and a change in contact type or duration. I link biological evaluation to change control, so each change gets a documented impact check, and I repeat chemical or biological work only where the change could alter patient exposure.”

8. What are common reasons for biocompatibility deficiencies?

Model answer: “Missing rationale for the endpoints chosen or omitted, testing a component or a non-final device, relying on a supplier statement such as USP Class VI as complete evidence, unclear sample preparation, no assessment of reprocessing for reusable devices, and no link between the biological conclusion and the risk file.”

Case study 1: a supplier changes an additive

Scenario: A fictional catheter company receives a notice that its resin supplier has changed a processing additive. The supplier says the resin still meets the same grade and no further testing is needed.

Ideal answer: “I would treat it as a change that needs an assessment, not accept the grade statement. I would ask for the additive identity and level, review the toxicological profile, and compare extractables from old and new material under the same conditions. If the chemistry shows no new or higher exposure, I document that and may not need further testing. If it does, I add the relevant endpoints. I would also update the supplier agreement to require advance notice of formulation changes.”

Case study 2: a cytotoxicity result is borderline

Scenario: At a fictional wound-care device company, the cytotoxicity test on the finished product shows a reactive result. The project lead asks you to repeat the test on a new lot.

Ideal answer: “I would not simply repeat it. I would check the laboratory report and sample preparation, then look for a cause: residual cleaning agent or sterilisation residue, a material or process change, or a laboratory issue. A repeat is reasonable only once the cause is understood, and a second pass does not cancel the first result. I would record both in the investigation, correct any process cause and reassess other endpoints that might be affected.”

Case study 3: a predicate’s material as the whole argument

Scenario: A fictional orthopedic company wants to skip biocompatibility testing because a predicate device uses the same polymer. The regulatory lead asks if that is enough.

Ideal answer: “Using the same polymer name alone is not enough. I would compare the full formulation, additives, manufacturing processes, surface finish, sterilisation, and the contact type and duration. If we can show the material and processing are equivalent, with documentation, that supports the argument. If there are differences, I would use chemical characterisation to bridge them or test the gaps. The plan should say which comparison we made and what we could not confirm.”

Mistakes to avoid

  • Describing biocompatibility as “running the ISO 10993 tests”.
  • Treating a supplier’s “ISO 10993 compliant” or “USP Class VI” statement as complete evidence for a finished device.
  • Testing raw material or components instead of the finished, sterilised device.
  • Repeating a failed test without investigating the cause.
  • Forgetting reprocessing for reusable devices, or ignoring change control.

Frequently asked questions

Do I need a toxicology degree?

Not always. Many roles are held by engineers or scientists who work with a toxicologist. A degree in toxicology or a related science helps for senior roles, and the toxicological risk assessment is usually written by a qualified toxicologist.

Is biocompatibility only a testing role?

No. Much of the work is planning, reviewing existing data, managing laboratories and suppliers, assessing changes and writing the rationale. Testing is only one source of evidence.

What other interview guides should I read?

Try the design controls and risk management guide, the sterilization and packaging guide and the supplier quality guide.

Model answers are examples; adapt them to your real experience. Companies and devices in the case studies are fictional. Standards and guidance are revised, so confirm the current edition against the official source.